TY - JOUR
T1 - Two-site evaluation of the Roche Elecsys Vitamin D total III assay
AU - Favresse, Julien
AU - Schiettecatte, Johan
AU - Wolff, Fleur
AU - Cotton, Frederic
AU - Elsen, Marc
AU - Eucher, Christine
AU - Rosseels, Catherine
AU - Anckaert, Ellen
N1 - Publisher Copyright:
© 2022 Walter de Gruyter GmbH, Berlin/Boston.
PY - 2022/9/1
Y1 - 2022/9/1
N2 - Objectives: The high request for vitamin D testing in the last decades has led manufacturers to develop assays on automated immunoassay platforms. The objective of this study was to evaluate the performance of the new Elecsys Vitamin D total III assay for the measurement of total 25(OH)D. Methods: A total of 844 serum samples collected in two clinical laboratories were used to evaluate the new Roche Elecsys Vitamin D total III assay. Comparisons with Roche Elecsys Vitamin D total II and liquid chromatography tandem mass spectrometry (LC-MS/MS) were carried out. Additionally, assay imprecision, linearity, matrix effects, biotin interference, cross-reactivity with 24,25(OH)2D3 and 3-epi-25(OH)D3, and outlier rate were evaluated for the Elecsys Vitamin D total III assay. Results: Only the comparison between LC-MS/MS and Roche Elecsys Vitamin D total III achieved the optimal specification for bias (i.e., <3.4%). Imprecision, linearity and matrix effects showed acceptable results. The biotin interference threshold was increased up to 1,200 ng/mL and the outlier rate was low (0.26%). The cross-reactivity with 24,25(OH)2D3 and 3-epi-25(OH)D3 was weak or modest in available patient samples. However, using SRM972a with a high level of 3-epi-25(OH)D3 (enriched) revealed an important cross-reactivity with both Roche Elecsys Vitamin D total II and III assays (+74.7% and +73.7%). Conclusions: In conclusion, the Roche Elecsys Vitamin D total III assay presents several advantages compared to the previous assay generation: higher biotin interference threshold, broader measuring range, and better comparability with LC-MS/MS. However, the cross-reactivity toward 3-epi-25(OH)D3 is still problematic in high titer samples.
AB - Objectives: The high request for vitamin D testing in the last decades has led manufacturers to develop assays on automated immunoassay platforms. The objective of this study was to evaluate the performance of the new Elecsys Vitamin D total III assay for the measurement of total 25(OH)D. Methods: A total of 844 serum samples collected in two clinical laboratories were used to evaluate the new Roche Elecsys Vitamin D total III assay. Comparisons with Roche Elecsys Vitamin D total II and liquid chromatography tandem mass spectrometry (LC-MS/MS) were carried out. Additionally, assay imprecision, linearity, matrix effects, biotin interference, cross-reactivity with 24,25(OH)2D3 and 3-epi-25(OH)D3, and outlier rate were evaluated for the Elecsys Vitamin D total III assay. Results: Only the comparison between LC-MS/MS and Roche Elecsys Vitamin D total III achieved the optimal specification for bias (i.e., <3.4%). Imprecision, linearity and matrix effects showed acceptable results. The biotin interference threshold was increased up to 1,200 ng/mL and the outlier rate was low (0.26%). The cross-reactivity with 24,25(OH)2D3 and 3-epi-25(OH)D3 was weak or modest in available patient samples. However, using SRM972a with a high level of 3-epi-25(OH)D3 (enriched) revealed an important cross-reactivity with both Roche Elecsys Vitamin D total II and III assays (+74.7% and +73.7%). Conclusions: In conclusion, the Roche Elecsys Vitamin D total III assay presents several advantages compared to the previous assay generation: higher biotin interference threshold, broader measuring range, and better comparability with LC-MS/MS. However, the cross-reactivity toward 3-epi-25(OH)D3 is still problematic in high titer samples.
KW - 25(OH)D
KW - immunoassay
KW - liquid chromatography tandem mass spectrometry (LC-MS/MS)
KW - vitamin D
UR - http://www.scopus.com/inward/record.url?scp=85135084338&partnerID=8YFLogxK
U2 - 10.1515/cclm-2022-0177
DO - 10.1515/cclm-2022-0177
M3 - Article
C2 - 35849629
AN - SCOPUS:85135084338
SN - 1434-6621
VL - 60
SP - 1598
EP - 1606
JO - Clinical Chemistry and Laboratory Medicine
JF - Clinical Chemistry and Laboratory Medicine
IS - 10
ER -